Model Number R SERIES
Device Problem Inaccurate Synchronization (1609)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/19/2023
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that during a cardioversion on a patient (age & gender unknown), the device's output time of sync r wave was incorrect.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
Zoll medical canada evaluated the device and the device performed to specification.The device passed the final test and was recertified and returned to the customer.Review of the returned video and ecg strip confirmed inconsistent r wave marking, and double counting are observed in the video due to the t wave having a tall/sharp peak, causing the device to detect it as a syncable r wave.The data records were of appropriate length to allow a satisfactory analysis of the ecg signal.However, if the device misplaces a sync marker over an incorrect wave due to a spiked t wave, the r series operator guide states, "verify that markers are clearly visible on the monitor and their location is appropriate and consistent from beat to beat.If necessary, use the lead and size buttons to establish settings that yield the most consistent sync marker pattern." it is important the user also verify that the sync marker is correct and adjust the lead view/ size in order to deliver the correct therapy.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key17120003
MDR Text Key317664594
Report Number1220908-2023-02198
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P160022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-