Zoll medical canada evaluated the device and the device performed to specification.The device passed the final test and was recertified and returned to the customer.Review of the returned video and ecg strip confirmed inconsistent r wave marking, and double counting are observed in the video due to the t wave having a tall/sharp peak, causing the device to detect it as a syncable r wave.The data records were of appropriate length to allow a satisfactory analysis of the ecg signal.However, if the device misplaces a sync marker over an incorrect wave due to a spiked t wave, the r series operator guide states, "verify that markers are clearly visible on the monitor and their location is appropriate and consistent from beat to beat.If necessary, use the lead and size buttons to establish settings that yield the most consistent sync marker pattern." it is important the user also verify that the sync marker is correct and adjust the lead view/ size in order to deliver the correct therapy.No trend is associated with reports of this type.
|