• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUROPACE, INC. NEUROPACE RNS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number RNS-320-K
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Headache (1880); Hematoma (1884)
Event Date 04/09/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).The rns neurostimulator and leads remain implanted and programmed for use.
 
Event Description
The patient had an open resection of the left anterior medial temporal lobe region on (b)(6) 2023 with revision of the prior placed bilateral hippocampal rns (unchanged on right, electrode moved to posterior margin of resection on the left).Rns device functioned normally after surgery.On (b)(6)/2023, after presenting with progressive headache, ct revealed a large left hemisphere subdural hematoma (sdh), the patient underwent reopening of the left hemisphere craniotomy to evacuate the hematoma.The sdh extended over much of the left hemisphere, including up to the anterior margin of the rns device location.Patient has been doing fine since the evacuation of hematoma procedure, no seizures, no symptoms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
5108822607
MDR Report Key17121230
MDR Text Key317111148
Report Number3004426659-2023-00027
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005267
UDI-Public010085554700526717200424
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 06/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberRNS-320-K
Device Catalogue Number1007694
Device Lot Number28164-1-1-1
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/19/2023
Initial Date FDA Received06/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age42 YR
Patient SexFemale
-
-