Model Number HL-90 |
Device Problem
Device Alarm System (1012)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Event Description
|
It was reported that the product has visible warning lights and audible sounds.Patient involvement is unknown.
|
|
Manufacturer Narrative
|
Other, other text: b3: date of event and d4:udi section are unknown, no product information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Other text: d10, h3, and h6 updated.Device evaluation: one device was returned for investigation in used condition.Functional testing of the device was done where water was put in the tank and the temperature check was attached and device turned on.The customer issue of visible lights with no audible sound was confirmed.The speaker on the printed circuit board (pcb) was not working.No action was taken due to the device being beyond economical repair.A device history review was not done due to the device being beyond a year from the manufacturing date.
|
|