BOSTON SCIENTIFIC CORPORATION WOLVERINE PERIPHERAL CUTTING BALLOON; Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring Back to Search Results
Model Number 39345-401510
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/19/2023
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: the device was returned for evaluation.A visual examination identified that the hypotube was severely kinked at more than one location.The hypotube was also completely separated at approximately 650mm distal of the strain relief.This type of damage is consistent with excessive force being applied to the device.No issues were identified with the shaft polymer extrusion of the device.A visual examination found that the balloon had been inflated.No issues or damage was noted to the balloon, blades, markerbands or tip of the device.All blades were present and fully bonded to the balloon material.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 29may2023.It was reported that the center part of the shaft got kinked.The 90% stenosed target lesion was located in the moderately tortuous and moderately calcified left popliteal artery.A 4.0mm x 15mm wolverine peripheral cutting balloon was selected for use.During the procedure, it was noted that the center part of the shaft got kinked.The procedure was completed with a different device.No patient complications were reported.However, device analysis revealed that a complete separation of the hypotube occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE PERIPHERAL CUTTING BALLOON
Type of Device
Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jeff wallner
4100 hamline ave n
arden hills, MN 55112
6515811560
MDR Report Key17122499
MDR Text Key317243063
Report Number2124215-2023-29783
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 06/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number39345-401510
Device Catalogue Number39345-401510
Device Lot Number0030114607
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/09/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-