OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problems Device Alarm System (1012); Display or Visual Feedback Problem (1184)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported, that the device displays red flashing lights.Patient involvement unknown.
 
Manufacturer Narrative
B3: date of event and d4:udi section is unknown.No product information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56, when additional reportable information becomes available.
 
Manufacturer Narrative
Email is: (b)(6).Evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection found a corroded drain fitting, cracked tank cover, broken led's, and damaged micro switch.Functional testing found the no disposable alarm was sounding and flashing; confirming the customer complaint.Root cause was attributed to a damaged micro switch.The arm is severed from the body of the switch.It was thought the broken microswitch lever happened from usage.The product's history records were reviewed and there were no non-conformance's nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17129940
MDR Text Key317240589
Report Number3012307300-2023-06244
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? No
Date Returned to Manufacturer06/06/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/30/2023
Initial Date FDA Received06/14/2023
Supplement Dates Manufacturer Received01/19/2024
Supplement Dates FDA Received02/15/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/13/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-