Model Number HL-390 |
Device Problems
Device Alarm System (1012); Display or Visual Feedback Problem (1184)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Event Description
|
It was reported, that the device displays red flashing lights.Patient involvement unknown.
|
|
Manufacturer Narrative
|
B3: date of event and d4:udi section is unknown.No product information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56, when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Email is: (b)(6).Evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection found a corroded drain fitting, cracked tank cover, broken led's, and damaged micro switch.Functional testing found the no disposable alarm was sounding and flashing; confirming the customer complaint.Root cause was attributed to a damaged micro switch.The arm is severed from the body of the switch.It was thought the broken microswitch lever happened from usage.The product's history records were reviewed and there were no non-conformance's nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
|
|