• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-N180
Device Problems Contamination /Decontamination Problem (2895); Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The customer suspected device was returned for investigation.The olympus service center confirmed the reported event during inspection and testing.Upon evaluation of the returned device the following defects were found, angel rubber glue separated, angel rubber glue discolored, light guide lens cracked, light guide lens chipped, insertion tube wrinkled near glue, bending angulations out of specifications, and connector corroded.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
A user facility reported to olympus that the evis exera ii gastrointestinal videoscope had no insufflation despite change of plug.Upon inspection and testing of the returned device, foreign material was found clogged in the scope channel due to insufficient cleaning.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.There was no report of patient harm or user injury associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.It is likely that the foreign material was of human origin.There have been no reports of deformation or damage to the air/water channels.The detection method is described in the olympus gif type n180 operation manual chapter 3 preparation and inspection preventive measures are described in the instruction manual olympus gif type n180 reprocessing manual chapter 5 reprocessing the endoscope.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17130909
MDR Text Key317350757
Report Number9610595-2023-08862
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04953170339080
UDI-Public04953170339080
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K100584
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-N180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/27/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-