OAKDALE LEVEL 1 HOTLINE LOW FLOW SYSTEM; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problem Device Alarm System (1012)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported that the unit is constantly alarming.Patient involvement is unknown.
 
Manufacturer Narrative
Other, other text: b3: date of event and d4:udi section are unknown, no product information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
One device was returned for evaluation with a cracked enclosure and tank cover, damaged power switch, printed circuit board (pcb), and their connection.Functional testing occurred and found that the "no disposable" alarm goes off, confirming the reported complaint.The root cause is a bent micro switch lever from repeated usage.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.A manufacturing dhr review was not performed because the device is beyond a year from its manufacture date and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE LOW FLOW SYSTEM
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17133741
MDR Text Key317211004
Report Number3012307300-2023-06262
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/30/2023
Initial Date FDA Received06/15/2023
Supplement Dates Manufacturer Received07/28/2023
Supplement Dates FDA Received08/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-