Other, other text: b3: date of event and d4:udi section are unknown, no product information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
One device was returned for evaluation with a cracked enclosure and tank cover, damaged power switch, printed circuit board (pcb), and their connection.Functional testing occurred and found that the "no disposable" alarm goes off, confirming the reported complaint.The root cause is a bent micro switch lever from repeated usage.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.A manufacturing dhr review was not performed because the device is beyond a year from its manufacture date and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.
|