Model Number 0998-00-0800-55 |
Device Problems
Premature Discharge of Battery (1057); Failure to Charge (1085); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/06/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) unit had a battery problem and it was thought to need maintenance.The iabp is connected to a patient, and always connect to the electricity (220v) the 2 batteries are in green, but once disconnected from the mains, one of the batteries exhausted in +/- 5 min.There was no patient harm reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the iabp and was unable to reproduce the reported issue.The fse performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released and cleared for clinical service.
|
|