The customer reported that the ceramic insulation at the distal tip of the inner resectoscope sheath was found to have a crack during reprocessing.The device was replaced, and the scheduled transurethral resection of the bladder tumor procedure was completed.There was a delay in replacing the equipment, but the amount of time is unknown.No death or injury and no impact to patient or other was reported.The device was inspected prior to use but no anomalies were noted.
|
The referenced device was returned for evaluation and the inspection confirmed the customer¿s reported issue.A large portion of the ceramic insulation at the distal tip of the device was broken off.No further information regarding the event was provided by the customer.A review of the (dhr) manufacturing and quality records for the affected lot number was reviewed without detecting any non-conformities or deviations regarding the described issue.The investigation is still in progress; however, if additional information becomes available, this report will be supplemented accordingly.Additional 510(k): k931994.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation (h6/h10).Based on the results of the investigation, the root cause could not be determined.It is likely the damage to the tip was caused by one of the following; a thermal/mechanically induced impact, wear and tear, improper handling, and/or a mechanical overload (such as impact from a fall, shock, or similar stress).It was further noted that it cannot be determined with certainty, whether there was pre-existing damage or if it had occurred during the last procedure or last reprocessing.The instructions for use (ifu) provides a warning that the ceramic tip can break due to mechanical loading or thermally induced straining: ¿4 before use warning infection control risk properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing inspecting the product visually inspect the product.Make sure that it has: -- no corrosion -- no dents -- no scratches ceramic insulation at distal end visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning risk of injury impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.¿ olympus will continue to monitor field performance for this device.
|