Catalog Number 03.168.011 |
Device Problem
Failure to Cut (2587)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/30/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: reporter is a j&j employee.H6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from depuy synthes reports an event in japan as follows: it was reported that on (b)(6)2023, the patient underwent orif surgery for right femoral neck fracture with fns drill bit.During the lateral locking screw insertion phase, it took time to reach the contralateral cortex when drilling.The surgeon pointed out that the drill bit was not sharp enough.Procedure was completed successfully with thirty(30) minutes of surgical delay.No further information is available.This report is for one (1) 4.3mm drill bit length 413mm.This is report 2 of 2 for complaint (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Part: 03.168.011, lot: f-33513, release to warehouse date: 04 may 2022, manufacturing site: werk selzach, expiration date: n/a, supplier: (b)(4).The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the device found blunting and rounding of cutting edges.It is not unreasonable that the condition identified in visual analysis would contribute to a dull condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed due to we were able to confirm dullness with damage observed, as this kind of evidence indicates constant and heavy usage of the device.The device exhibits damage consistent with repeated use.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the [drill bit ø4.3 l413] would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
|