Catalog Number SEPX-8-6-40-135
Device Problems Deformation Due to Compressive Stress (2889); Activation, Positioning or Separation Problem (2906); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/26/2023
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Physician intended to use a protégé rx carotid stent system with a 5mm spider fx embolic protection device during treatment of a plaque lesion in the patient¿s mid common carotid artery.Moderate vessel calcification and moderate vessel tortuosity are reported.Lesion exhibited 85% stenosis.There was no damage noted to packaging.There were no issues noted when removing the device from the hoop/tray.Device prepped per the ifu with no issues identified.The lesion was not pre-dilated.The thumbscrew/lock-pin was checked for securement prior to procedure and the thumbscrew/lock-pin was removed prior to attempted deployment.No resistance was encountered during delivery to the lesion and no excessive force was used.Stent deformation in vivo during positioning/ deployment is reported.During the surgery, the distal end of the stent was kinked and bent before it was deployed and could not continue to used, so it was withdrawn from the patient immediately.No intervention was required for removal of the device and the device was safely removed from the patient.No patient injury reported.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGE RX CAROTID STENT SYSTEM
Type of Device
STENT, CAROTID
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17141840
MDR Text Key317307598
Report Number2183870-2023-00223
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P060001 EXPO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSEPX-8-6-40-135
Device Lot NumberB524256
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/29/2023
Date Device Manufactured02/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-