Model Number 0998-00-0800-53 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/07/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) , the cardiosave intra-aortic balloon pump (iabp) units 9v critical alarm battery failed prematurely.There was no patient involvement.
|
|
Manufacturer Narrative
|
Additional point of contact: name: (b)(6).Phone number: (b)(6).Email id: (b)(6).Contact department: clinical engineering.Contact person occupation: biomedical engineering.A getinge field service engineer (fse) during a preventative maintenance (pm) service repair found that the 9v critical alarm battery had failed prematurely. the battery showed signs of corrosion around the seams of the battery.To fix this issue fse replaced the battery and completed a full pm.All tests passed to factory specifications.Unit returned to the customer and released for clinical use.
|
|
Event Description
|
N/a.
|
|