LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00Z
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/16/2023
Event Type  malfunction  
Manufacturer Narrative
H10: livanova deutschland manufactures the s5 double head pump.The incident occurred in haifa, israel.The unit was tested by the customer and found that pump head complete is defective.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a s5 double head pump gave a motor control failure error message during priming.There was no patient involvement.
 
Manufacturer Narrative
H10: the involved unit has been sent to the manufacturer site for repair and a replacement pump has been provided to the customer.Based on the technical intervention, most likely the root cause of the reported event was an isolated early failure of the pump head.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17150168
MDR Text Key317775653
Report Number9611109-2023-00270
Device Sequence Number1
Product Code DWB
UDI-Device Identifier04033817902737
UDI-Public(01)04033817902737(11)230209
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00Z
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/28/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/09/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-