Model Number 10-85-00Z |
Device Problem
Pumping Stopped (1503)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/16/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: livanova deutschland manufactures the s5 double head pump.The incident occurred in haifa, israel.The unit was tested by the customer and found that pump head complete is defective.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a s5 double head pump gave a motor control failure error message during priming.There was no patient involvement.
|
|
Manufacturer Narrative
|
H10: the involved unit has been sent to the manufacturer site for repair and a replacement pump has been provided to the customer.Based on the technical intervention, most likely the root cause of the reported event was an isolated early failure of the pump head.
|
|
Event Description
|
See initial report.
|
|
Search Alerts/Recalls
|