SHIRAKAWA OLYMPUS CO., LTD. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problem Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/20/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to an olympus repair facility, and an evaluation of the device was performed.Upon inspection and testing of the unit, error code e102 was displayed, caused by a defective converter.In addition, corrosion was found on the flat cable.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
A customer reported error code e102, the xenon lamp not illuminating, and the light source in emergency lamp mode on the visera elite xenon light source during a laparascopic procedure.The procedure was completed using another device.There were no reports of patient harm.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the e102 could not be identified, nor could the phenomenon be confirmed/reproduced.Also, based on the results of the legal manufacturer's investigation, a definitive root cause of the e103 and light source switching to emergency lamp mode could not be identified.However, it¿s likely the cause is related to a faulty power supply.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17154840
MDR Text Key318082970
Report Number3002808148-2023-06130
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170301995
UDI-Public4953170301995
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 08/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/22/2023
Initial Date FDA Received06/19/2023
Supplement Dates Manufacturer Received07/14/2023
Supplement Dates FDA Received08/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/05/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-