BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C1009K
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/25/2023
Event Type  malfunction  
Event Description
It was reported that the bladder of a large volume infusor ruptured.The bladder rupture occurred while filling the device prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H4: the lot was manufactured between april 26, 2022 to april 27, 2022.H10: the device was received for evaluation.A visual inspection with the naked eye was performed and noted the bladder has been ruptured.The ruptured bladder was examined for signs of abnormalities that could have contributed to the rupture.There were no signs of abnormalities.The reported condition was verified.The cause of the condition could not be determined; however, a possible cause was due to inherent variation in the material, as bladders are manufactured with rubber.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17157857
MDR Text Key318099014
Report Number1416980-2023-03005
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412022529
UDI-Public(01)00085412022529
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C1009K
Device Lot Number22D023
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/25/2023
Initial Date FDA Received06/19/2023
Supplement Dates Manufacturer Received07/11/2023
Supplement Dates FDA Received07/17/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
N/A.
-
-