H4: the lot was manufactured between april 26, 2022 to april 27, 2022.H10: the device was received for evaluation.A visual inspection with the naked eye was performed and noted the bladder has been ruptured.The ruptured bladder was examined for signs of abnormalities that could have contributed to the rupture.There were no signs of abnormalities.The reported condition was verified.The cause of the condition could not be determined; however, a possible cause was due to inherent variation in the material, as bladders are manufactured with rubber.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|