This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was returned to olympus.Device evaluation confirmed the cold light connector is missing.Additional defects found during device evaluation: the direction of view is out of order (grey edge in the field of view), and dirt visible in eyepiece window.The serial number for the affected device is not known in any of olympus's systems.Therefore, a device history record review could not be performed.However, olympus only releases products to the market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, it is likely that the event occurred due to excessive use and/or unsuited/inappropriate handling and/or the apply of external force (e.G.By fall, impact or shock).Olympus will continue to monitor field performance for this device.
|