OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Leak/Splash (1354)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/18/2023
Event Type  malfunction  
Manufacturer Narrative
Date of event and d4: udi number is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the device was leaking.Patient involvement unknown.
 
Manufacturer Narrative
Device available for evaluation; h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was received with a cracked water tank cover, corroded quick connect, front cover had scratches on it, display label was dented, and reflux plug was discolored.Functional testing confirmed the complaint, and the root cause was due to a cracked tank cover.However, it was unknown what caused the crack.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced water tank cover, quick connect, front cover, reflux plug and microswitch.Device passed all functional and delivery tests.
 
Event Description
Additional information was received confirming there was no injury and no patient involvement.
 
Manufacturer Narrative
Other text: additional information: b3, b5, and h6 - health effects codes, corrected data: a1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17162182
MDR Text Key317490184
Report Number3012307300-2023-06315
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/02/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-