Model Number UHI-4
Device Problems Power Problem (3010); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/25/2023
Event Type  malfunction  
Event Description
The customer reported to olympus, during preparation for a therapeutic gynecological endometrial polyp procedure, the high flow insufflation unit exhibited alarm sound and the front panel turned off.It was noted that the intended procedure was completed.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
The subject device has been returned to olympus for evaluation and investigation.During evaluation, it was observed, visual inspection found no abnormalities, confirmation of actual equipment confirmed that the alarm sounds and the front-panel light was off.A review of the device history record (dhr) found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the problem occurred when the sensor on the main circuit board failed.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17162982
MDR Text Key318074396
Report Number3002808148-2023-06163
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 06/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-