MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 11 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 11 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number 42-5221-003-11
Device Problem Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/30/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4) d10: medical product: tibia cemented 5 degree stemmed right size c: catalog#42532006402, lot#65481553.G2: foreign: japan customer has indicated that the product is in process of being returned to zimmer biomet for evaluation.The investigation is in progress.Upon completion of the investigation, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient underwent an initial total knee arthroplasty.During the procedure, the surgeon was unable to mate the articular surface with the tibial plate despite multiple attempts.There was a surgical delay of less than thirty (30) minutes however no patient impact or adverse events have been reported as a result of the malfunction.The procedure was successfully completed with a second articular surface without further complication.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; d9; g1; g3; g6; h1; h2; h3; h6.Visual inspection of the return product identified the distal surface to exhibit damage with the locking features flared out.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 11 MM HEIGHT
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17164826
MDR Text Key317825421
Report Number3007963827-2023-00170
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K150090
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number42-5221-003-11
Device Catalogue Number42522100311
Device Lot Number65421535
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/30/2023
Initial Date FDA Received06/20/2023
Supplement Dates Manufacturer Received07/19/2023
Supplement Dates FDA Received07/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient SexPrefer Not To Disclose
-
-