It was reported that log files revealed a few s3 system fault alarms on 28jan2023 at 17:52 and on 31jan2023 at 9:20 and 9:21 that occurred after the console was powered on.These alarms correlated to a ¿can bus send error¿ sub-fault, and the console was not observed to have been in patient use during these times.The console was observed to have been shut down shortly after each alarm, resolving the alarms.
|
Manufacturer's investigation conclusion: the reported event of s3: system fault alarms was confirmed via the log file.A log file was extracted from the centrimag console when it was returned and evaluated under a separate event for a separate reason.A few s3: system fault alarms were observed on (b)(6) 2023 and on (b)(6) 2023, occurring shortly after the console was powered on.These alarms correlated to a sub-fault that was indicative of a communication error, and the console was not observed to have been in patient use during these times.The console was observed to have been shut down shortly after each alarm.The centrimag console (serial number (b)(6)) was not returned for analysis for this event.The root cause of the reported event was unable to be conclusively determined through this analysis.Review of the device history record for centrimag console, serial number (b)(6), showed the device was manufactured in accordance with manufacturing and qa specifications.The 2nd generation centrimag system operating manual section 3 "warnings and precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 11.1-"appendix i ¿ console alarms and alerts" contains a list of console alarms and alerts, including s3 alarms, as well as appropriate operator response to these events.No further information was provided.The manufacturer is closing the file on this event.
|