• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNOCH DUO FLUID CART; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH DUO FLUID CART; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Device Problem Suction Problem (2170)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/30/2023
Event Type  malfunction  
Event Description
It was reported that there was a suction issue.The event timing was outside of surgery.There is no harm or delay reported.No adverse events were reported as a result of this malfunction due diligence is complete and no additional information is available.Upon investigation, there was a char in the fuse.
 
Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record identified the following related repair: there was char in the fuse.A fuse was blown and was replaced which resolved the reported issue.Root cause - the vacuum pump is required to create suction in order for the unit to function as intended.Mechanical failure can occur within the vacuum pump assembly with either the pump failing to generate negative pressure, or with a failure of the carbon filter.Over time and with continuous use, the internal integrity of the vacuum pump can degrade and no longer be able to create the pressure needed for desired suction levels.Likewise, the carbon filter can back out of the assembly or over tighten, again causing suction failure.A mechanical failure of the component will cause an overall suction failure in the unit and will prompt vacuum sensor errors.Electrical failure can occur when the unit draws an excess of electrical current; this will cause the fuse for the vacuum pump to trip, denying the pump power and causing the component to fail and prompt vacuum sensor errors.The event was confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key17184152
MDR Text Key317946408
Report Number0001954182-2023-00074
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00889024465992
UDI-Public(01)00889024465992
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00514010100
Device Lot Number0029623
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/28/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-