Model Number HL-90 |
Device Problem
Defective Component (2292)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Event Description
|
It was reported, that the device motor is broken.There has been no report of patient involvement, or no observable clinical symptoms, or a change in symptoms identified in the patient.
|
|
Manufacturer Narrative
|
Udi section of d4 is unknown.No product information has been provided to date.B3: date of event is unknown.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56, when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
One device was received for investigation with a chipped front cover, cracked water tank cover, j201 pin missing, line cord has a crack showing wires, quick connect corroded and a crack under quick connect.The device was functionally tested and the reported problem was duplicated.The complaint is confirmed.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
|
|