Model Number RD900
Device Problems Break (1069); Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A healthcare facility in belgium reported that the gas inlet port of a rd900 neopuff infant resuscitator was damaged.There was no reported patient involvement.
 
Manufacturer Narrative
(b)(4).Fisher & paykel healthcare (f&p) is in the process of completing our investigation.We will provide a follow-up report upon completion of our investigation.
 
Manufacturer Narrative
(b)(4).Method: the rd900 neopuff infant resuscitator was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the information and photography provided by the customer and our knowledge of our product.Results: visual inspection of the photograph provided by the customer showed the the inlet port of the rd900 neopuff infant resuscitator was cracked.Conclusion: without the complaint device, we are unable to determine the cause of the reported fault.The neopuff is a portable, reusable device used to assist in the delivery of respiratory breaths to infants until adequate spontaneous breathing occurs.Being a portable device, the neopuff can be susceptible to impact damage, for instance when accidentally dropped or subjected to considerable external force.The neopuff technical manual warns against dropping the neopuff or subjecting it to impact damage which may cause the unit to operate incorrectly.If the neopuff is suspected to have been damaged, the manometer and valve system should be performance tested.In addition, the neopuff user instructions state that the user should "check manometer reads zero with no gas flow" and check the pressure settings "prior to every use of the neopuff".As mentioned, the neopuff technical manual states the following: - dropping the neopuff / perivent infant resuscitator or other similar forms of impact may cause damage resulting in incorrect operation of the unit.If you suspect damage to have occurred, please perform checks as outlined [in the manual] before connection to a patient.Each neopuff unit is assembled and 100% tested in the production line to verify that each unit conforms to critical product specifications.Any unit that fails is rejected.The subject neopuff would have met the requirements at the time of production.
 
Event Description
A healthcare facility in belgium reported that the gas inlet port of a rd900 neopuff infant resuscitator was damaged.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOPUFF INFANT RESUSCITATOR
Type of Device
BTL
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key17188054
MDR Text Key317994906
Report Number9611451-2023-00550
Device Sequence Number1
Product Code BTL
UDI-Device Identifier09420012410948
UDI-Public(01)09420012410948
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K892885
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRD900
Device Catalogue NumberRD900
Device Lot Number081010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/24/2023
Initial Date FDA Received06/22/2023
Supplement Dates Manufacturer Received08/10/2023
Supplement Dates FDA Received09/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/10/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-