• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number ATG80144
Device Problems Detachment of Device or Device Component (2907); Migration (4003)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/13/2023
Event Type  malfunction  
Event Description
It was reported during an angioplasty procedure, balloon allegedly separated from the shaft.It was further reported that the balloon allegedly started migrating thus it was stented in place.The current status of the patient was unknown.
 
Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiration date: 02/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported balloon detachment could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiration date: 02/2026), g3, h6 (method).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported during an angioplasty procedure, the pta balloon allegedly separated from the shaft and started migrating.Reportedly, it had to be stented in place.There was no reported patient injury.
 
Event Description
It was reported during an angioplasty procedure, balloon allegedly separated from the shaft.It was further reported that the balloon allegedly started migrating thus it was stented in place.There was no reported patient injury.
 
Manufacturer Narrative
Additional information was received, and the file was reassessed for reportability and determined to be reportable as malfunction.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.D4 (expiration date: 02/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD
Type of Device
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17189994
MDR Text Key317730810
Report Number2020394-2023-00443
Device Sequence Number1
Product Code DQY
UDI-Device Identifier00801741060755
UDI-Public(01)00801741060755
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 09/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberATG80144
Device Catalogue NumberATG80144
Device Lot Number93PH0094
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/31/2023
Initial Date FDA Received06/23/2023
Supplement Dates Manufacturer Received07/13/2023
09/07/2023
Supplement Dates FDA Received08/09/2023
09/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-