Model Number 71992-01 |
Device Problem
High Readings (2459)
|
Patient Problem
Diabetic Ketoacidosis (2364)
|
Event Date 02/01/2023 |
Event Type
Injury
|
Event Description
|
A high reading issue was reported with the adc device.A customer reported receiving an unspecified high scan on the sensor however, no symptoms were reported.The customer had contact with a healthcare professional who obtained a laboratory result of 500 mg/dl and diagnosed the customer with diabetic ketoacidosis (dka) and was provided antibiotics.No further information was provided.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A high reading issue was reported with the adc device.A customer reported receiving an unspecified high scan on the sensor however, no symptoms were reported.The customer had contact with a healthcare professional who obtained a laboratory result of 500 mg/dl and diagnosed the customer with diabetic ketoacidosis (dka) and was provided antibiotics.No further information was provided.There was no report of death or permanent impairment associated with this event.
|
|
Search Alerts/Recalls
|