• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC KINCISE¿ AUTOMATED SURGICAL IMPACTOR; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC KINCISE¿ AUTOMATED SURGICAL IMPACTOR; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Model Number 1000-00-101
Device Problems Material Fragmentation (1261); Unintended System Motion (1430); Device Slipped (1584)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2023
Event Type  malfunction  
Event Description
It was reported that during an unspecified surgical procedure, it was discovered the impactor device was broken in the center, at the connection point between the base and the barrel, causing the device to separate and continuously impact without pressing the trigger.It was reported that there were no delays to the surgical procedure.It was not reported if a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.E1: the initial reporter's facility address and phone number were not provided.As of this date, the device has not been returned for evaluation.Therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, it was determined that the reported condition that the device was broken causing the device to separate and continuously impact without pressing the trigger was not confirmed.Therefore, an assignable root cause for the reported condition of unintended activation/motion and device fractured was not determined.However, during evaluation, it was determined that the device would not run and was loose.It was further determined that the device failed pretest for visual assessment, impactor operation assessment and final assessment.The assignable root cause was determined to be due to component failure from wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KINCISE¿ AUTOMATED SURGICAL IMPACTOR
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
TECHTRONIC INDUSTRIES MEDICAL, INC
1428 pearman dairy rd
anderson SC
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 
3035526892
MDR Report Key17197290
MDR Text Key317820279
Report Number1045834-2023-00970
Device Sequence Number1
Product Code GEY
UDI-Device Identifier00845384018520
UDI-Public00845384018520
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1000-00-101
Device Catalogue Number1000-00-101
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/15/2023
Initial Date FDA Received06/26/2023
Supplement Dates Manufacturer Received06/28/2023
Supplement Dates FDA Received07/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-