Lot Number 0030953223
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Air Embolism (1697); Non specific EKG/ECG Changes (1817)
Event Date 06/12/2023
Event Type  Injury  
Manufacturer Narrative
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
During a polarx case a polarsheath was selected for use.There was a huge transient st elevation only during the first seconds of ablation on left superior pulmonary vein (lspv).The patient was under general anesthesia.After the preparation of the polarsheath; flushed with a syringe of 20 cc of saline water with the edge higher than the handle), the dilator was slowly removed and they let the blood flow out and at the same time they tap the 3-way with a syringe.Catheter preparation include inflation and air bubble removal in a bowl with saline water.The embolism was resolved in some minutes by providing high volume of oxygen.The console had a strange behavior linked to valve sv8: the balloon can be inflated only in slow mode.There were no errors that occurred during the case.The patient fully recovered and was discharged from the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARSHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
302 parkway
global park, la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key17201500
MDR Text Key317861190
Report Number2124215-2023-32675
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeSZ
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/11/2023
Device Lot Number0030953223
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/12/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/30/2023
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-