MEDOS INTERNATIONAL SÃ RL CH 5.5 TI CORT FIX 6X45MM; ORTHOSIS, SPONDYLOLISTHESIS SPINAL FIXATION
|
Back to Search Results |
|
Model Number 186731645 |
Device Problem
Device Slipped (1584)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 01/01/2023 |
Event Type
Injury
|
Event Description
|
It was reported that on an unknown date there was rod translation due to a loose set screw.A complaint was submitted at the time.The patient then required revision surgery, where the surgeon stated that the rod had migrated.They were only able to recover the pedicle screw.The surgery was completed successfully.There was no patient involvement.Patient status or outcome required revision surgery.Concomitant device reported: unk - rods: spine (part# unknown, lot# unknown, quantity# unknown).This report is for one (1) 5.5 ti cort fix 6x45mm.This is report 1 of 2 for complaint (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D2b: additional product code: kwp,mni,nkb and kwq.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j sales representative.H6: a manufacturing record evaluation was performed for the finished device.Product code : 186731645, lot number# 324965, it was electronically reviewed and the following non-conformance has been observed: jbl-nr0012400, no nonconformances /manufacturing irregularities related to the malfunction were identified.The product was released on: 25/11/2021.Manufacturing site: jabil le locle.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
This complaint is related to (b)(4).
|
|
Manufacturer Narrative
|
Device returned.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Event description updated.Implant date added.510k added.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that 5.5 ti cort fix 6x45mm presented no visual damage or defects that could contributed to a product malfunction.A functional test was not able to be performed since the rod was not returned for evaluation.A dimensional inspection was not performed for the 5.5 ti cort fix 6x45mm since it was not applicable.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the observed condition of the 5.5 ti cort fix 6x45mm would not contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|