Based on the investigation analysis, the reported event at 3:46 pm est on may 28, 2023, could not be confirmed due to no corresponding calibration entry at the time.However, prior to the reported event, there was a calibration entry at 5:28 pm est on (b)(6) 2023, which showed good agreement with the sensor reading at that time.Following the reported event, the sensor performance was evaluated, and the sensor was performing within expectations.The user did not seek any medical attention and self-treated by eating candy.The current system performance is within expectation and as per dms, the user is currently using the system with up-to-date information.H3.Device evaluated by manufacturer? yes.H6.Investigation findings updated to 213.H6.Investigation conclusions updated to 67.
|