SENSEONICS INC. EVERSENSE TRANSMITTER; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102208-600
Device Problem Imprecision (1307)
Patient Problem Hypoglycemia (1912)
Event Date 05/28/2023
Event Type  Injury  
Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Event Description
On (b)(6) 2023, senseonics was made aware of an adverse event where the user received a false hypoglycemia alert due to a sensor inaccuracy.
 
Manufacturer Narrative
Based on the investigation analysis, the reported event at 3:46 pm est on may 28, 2023, could not be confirmed due to no corresponding calibration entry at the time.However, prior to the reported event, there was a calibration entry at 5:28 pm est on (b)(6) 2023, which showed good agreement with the sensor reading at that time.Following the reported event, the sensor performance was evaluated, and the sensor was performing within expectations.The user did not seek any medical attention and self-treated by eating candy.The current system performance is within expectation and as per dms, the user is currently using the system with up-to-date information.H3.Device evaluated by manufacturer? yes.H6.Investigation findings updated to 213.H6.Investigation conclusions updated to 67.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE TRANSMITTER
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20875-7005
MDR Report Key17206226
MDR Text Key317901030
Report Number3009862700-2023-00141
Device Sequence Number1
Product Code QHJ
UDI-Device Identifier00817491022912
UDI-Public817491022912
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date10/24/2022
Device Model Number102208-600
Device Catalogue NumberFG-5900-01-001
Device Lot Number133067
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/24/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age72 YR
Patient SexFemale
-
-