Model Number N/A
Device Problem Improper Chemical Reaction (2952)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2023
Event Type  malfunction  
Event Description
It was reported that during surgery the cement paste set too fast.The cement was taken from the refrigerator about (10) minutes before the start of the surgery.The surgeon began using the cement approximately (4) minutes after the start of mixing.Insertion of the implant was about (5) minutes from the start of mixing.The cement then hardened in about (7) minutes.No impact on a patient.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Manufacturer Narrative
(b)(4).G2 ¿ foreign ¿ "japan".No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A retain sample of batch has been tested in the laboratory under standardized conditions.No unusual behavior during mixing, handling or setting.The reported behavior of the cement cannot be reproduced.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Device used for treatment.Medical records were not provided.With the available information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIPAC 80 REFOB BONE CMT R-3
Type of Device
BONE CEMENT
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
Manufacturer (Section G)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR   26903
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key17209551
MDR Text Key318203902
Report Number3006946279-2023-00055
Device Sequence Number1
Product Code MBB
UDI-Device Identifier04040029922385
UDI-Public(01)04040029922385(17)230630(10)AY47CI1605
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2023
Device Model NumberN/A
Device Catalogue Number4712500398-3
Device Lot NumberAY47CI1605
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-