(b)(4).G2 ¿ foreign ¿ "japan".No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A retain sample of batch has been tested in the laboratory under standardized conditions.No unusual behavior during mixing, handling or setting.The reported behavior of the cement cannot be reproduced.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Device used for treatment.Medical records were not provided.With the available information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|