• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION STOPCOCK; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION STOPCOCK; STOPCOCK, I.V. SET Back to Search Results
Catalog Number ZMC7401
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/31/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that five (5) stopcocks had small holes which lead to a leak; a blood leak was noted.The issue occurred during patient infusion.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the actual sample and a photograph of the sample was received for evaluation.Visual inspection noted cracks along the frame leading to the cap.A video sample was provided which indicated that the unit was leaking from the position of the visually identified cracks.The reported condition was verified.The cause of the condition was determined to be a manufacturing related issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STOPCOCK
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MALTA
a47 industrial estate
marsa
MT  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17209728
MDR Text Key318104661
Report Number1416980-2023-03174
Device Sequence Number1
Product Code FMG
UDI-Device Identifier05413760157153
UDI-Public(01)05413760157153
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup
Report Date 07/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberZMC7401
Device Lot Number21G12V166
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 05/31/2023
Initial Date FDA Received06/27/2023
Supplement Dates Manufacturer Received07/11/2023
Supplement Dates FDA Received07/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NI.
-
-