MAUDE Adverse Event Report: SYNTHES GMBH ACCESS KIT 10 G DIAM TIP SIDE-OPEN DOUBL; CEMENT, BONE, VERTEBROPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH ACCESS KIT 10 G DIAM TIP SIDE-OPEN DOUBL; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Catalog Number 03.804.514S
Device Problems Problem with Sterilization (1596); Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/07/2023
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j employee.G4: device is not distributed in the united states, but is similar to device marketed in the usa.H3, h4, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot part# 03.804.514s lot#22e19-1 manufacturing site: werk selzach logistik supplier: sfm medical devices gmbh release to warehouse date: 19 may 2022 expiration date: 19 may 2027 a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in japan as follows: it was reported that during a percutaneous vertebroplasty procedure on (b)(6)2023, when the syringe and the cement needle were taken out of the kit, the surgeon said that he could not connect the two because something was stuck in the cement needle.A nurse tried to connect them, and she couldn¿t either for the same reason.The nurse was unable to remove the substance stuck in the cement needle using tweezers.A new kit was opened, and the procedure was completed successfully with a 30-minute delay.There was no adverse consequence to the patient.This report involves one access kit 10 g diam tip side-open doubl.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCESS KIT 10 G DIAM TIP SIDE-OPEN DOUBL
Type of Device
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK SELZACH
bohackerweg 5
selzach 2425
SZ   2425
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key17209761
MDR Text Key318118397
Report Number8030965-2023-08040
Device Sequence Number1
Product Code NDN
UDI-Device Identifier07611819525375
UDI-Public(01)07611819525375
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.804.514S
Device Lot Number22E19-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/06/2023
Initial Date FDA Received06/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/19/2022
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-