Model Number G247 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Unspecified Infection (1930); Fluid Discharge (2686)
|
Event Date 06/19/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was part of a system revision due to infection.The physician opened the pocket and brown fluid was aspirated and was sent for cultures.There were no additional adverse patient effects reported.This crt-d was explanted.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, additional information from product investigation indicates that the allegation against the device was not confirmed.The device was analyzed, passed all the baseline tests and exhibited normal device functions.This supplemental is being filed to capture d9: returned to manufacturer date and investigation results.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was part of a system revision due to infection.The physician opened the pocket and brown fluid was aspirated and was sent for cultures.There were no additional adverse patient effects reported.This crt-d was explanted.
|
|
Search Alerts/Recalls
|