• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS INC. IMPRESS®.BRAIDED CATHETER; PERIPHERAL CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS INC. IMPRESS®.BRAIDED CATHETER; PERIPHERAL CATHETERS Back to Search Results
Catalog Number 46535RIM/B
Device Problems Fracture (1260); Material Separation (1562)
Patient Problem Foreign Body In Patient (2687)
Event Date 06/22/2023
Event Type  Injury  
Event Description
The account alleges that during an angiogram procedure the tip of the catheter detached within the patient.While attempting to cross through the common iliac artery, the tip of the device got stuck within the highly calcified vessel and detached within the patient.There was an attempt to remove the foreign body using a snare, and was unsuccessful.The patient was sent to surgery to remove the detached tip.No additional patient consequence to report.
 
Manufacturer Narrative
The suspect medical device will not be returning for engineering evaluation.A lot number was reported, and a review of the manufacturing history record is in progress.
 
Manufacturer Narrative
The suspect medical device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.Should the device be returned later, the investigation will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS®.BRAIDED CATHETER
Type of Device
PERIPHERAL CATHETERS
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS INC.
14646 kirby drive
houston TX 77047
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS INC.
14646 kirby drive
houston TX 77047
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
MDR Report Key17213892
MDR Text Key317985701
Report Number3011642792-2023-00046
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00884450668122
UDI-Public884450668122
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K053171
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number46535RIM/B
Device Lot NumberI2523327
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received07/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/18/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-