Catalog Number 46535RIM/B |
Device Problems
Fracture (1260); Material Separation (1562)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 06/22/2023 |
Event Type
Injury
|
Event Description
|
The account alleges that during an angiogram procedure the tip of the catheter detached within the patient.While attempting to cross through the common iliac artery, the tip of the device got stuck within the highly calcified vessel and detached within the patient.There was an attempt to remove the foreign body using a snare, and was unsuccessful.The patient was sent to surgery to remove the detached tip.No additional patient consequence to report.
|
|
Manufacturer Narrative
|
The suspect medical device will not be returning for engineering evaluation.A lot number was reported, and a review of the manufacturing history record is in progress.
|
|
Manufacturer Narrative
|
The suspect medical device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.Should the device be returned later, the investigation will be reopened.
|
|
Search Alerts/Recalls
|