• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN917285
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2023
Event Type  malfunction  
Event Description
It was reported that after the patient had been on the pump for 45 days, the pump started "making audible clicking noise when pumping with changes to bpw".Troubleshooting included 2 manual purges and emptying of the condensation bottle which resolved the problem but only for approximately 15-20 seconds, then the clicking resumed.As a result, the pump was exchanged for another pump.No patient harm or injury.The patient status is reported as "fine".
 
Manufacturer Narrative
(b)(4).Other remarks: n/a.Corrected data: n/a.
 
Manufacturer Narrative
(b)(4).The reported complaint of "changes to bpw" was not able to be confirmed as the iabp part was not returned for investigation.According to the attached service report, the issue was not able to be duplicated; however, the augmentation valve was replaced.A device history record (dhr) review was conducted for the lot number with no relevant findings.Based on dhr review, the product met specification upon release.The root cause of the complaint was undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.
 
Event Description
It was reported that after the patient had been on the pump for 45 days, the pump started "making audible clicking noise when pumping with changes to bpw".Troubleshooting included 2 manual purges and emptying of the condensation bottle which resolved the problem but only for approximately 15-20 seconds, then the clicking resumed.As a result, the pump was exchanged for another pump.No patient harm or injury.The patient status is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 OPTIMUS IABP NA/EMEA
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville 27560
MDR Report Key17215961
MDR Text Key318022063
Report Number3010532612-2023-00347
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10801902172065
UDI-Public10801902172065
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN917285
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/01/2023
Initial Date FDA Received06/28/2023
Supplement Dates Manufacturer Received07/21/2023
Supplement Dates FDA Received07/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient SexMale
-
-