Model Number 6662
Device Problem Wireless Communication Problem (3283)
Patient Problems Failure of Implant (1924); Inadequate Pain Relief (2388); Movement Disorder (4412)
Event Date 06/07/2023
Event Type  malfunction  
Manufacturer Narrative
Exact date of event is unknown.
 
Event Description
It was reported that the patient was unable to connect external devices to their ipg following an unrelated procedure.Troubleshooting with a company representative was successful in connecting to the ipg to the patient's programmer.The device is providing therapy.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Manufacturer Narrative
H6: health effect - clinical code: should not have been 2388 - inadequate pain relief, the correct code is 4412 - movement disorder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR
Type of Device
DBS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key17221099
MDR Text Key318050539
Report Number1627487-2023-02875
Device Sequence Number1
Product Code MHY
UDI-Device Identifier05415067020260
UDI-Public05415067020260
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/04/2021
Device Model Number6662
Device Catalogue Number6662
Device Lot Number6865592
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/05/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/06/02/2017/001-C
Patient Sequence Number1
Treatment
DBS BURR HOLE CAP; DBS EXTENSION (X2); DBS LEAD (X2)
Patient SexMale
Patient Weight68 KG
-
-