It was reported that the bd posiflush¿ syringe was damaged.The following information was provided by the initial reporter, translated from chinese to english: on (b)(6), 2022, in the treatment room of the department, after the patient's intravenous infusion is completed, the tube is sealed and the pre filled catheter irrigator is opened.The outside of the syringe was found broke,replace syringe immediately received an update from the sales representative, and the description of the incident was updated as follows: the incident has been communicated with relevant customers of omit the outer packaging of the product is intact and undamaged.When the packaging was opened to push the core rod to activate the stop ring device, it was found that the integrity of the two wings at the end of the syringe was damaged.For fear of possible risks, a new one was used.
|
H.6.Investigation summary: a device history record review was completed by our quality engineer team for provided material number 306594 and lot number 1322102.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.H3 other text: see h.10.
|