Model Number 71992-01 |
Device Problem
Low Readings (2460)
|
Patient Problems
Dehydration (1807); Hyperglycemia (1905); Nausea (1970); Loss of consciousness (2418); Confusion/ Disorientation (2553); Convulsion/Seizure (4406)
|
Event Date 05/28/2023 |
Event Type
Injury
|
Event Description
|
A low readings issue was reported with the abbott diabetes care (adc) device.A customer reported receiving an unspecified low scan on the sensor when compared to an unspecified result obtained from a competitor meter.The customer experienced symptoms described as "disorientation, sticky lips, thirst, nausea, seizure, and a loss of consciousness", and was unable to self-treat.The customer had contact with a healthcare professional who administered drip administration with saline for the diagnosis of hyperglycemia.No further information was reported.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
The customer¿s product has been requested for investigation.A follow-up report will be filed once additional information is obtained.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.The sensor plug is fully seated and no issues were observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Inspected the plug assembly, no issues were observed.The current was applied to the sensor to perform accuracy testing while in the test fixture.All results were within specification.Poise voltage and sensor thermistor testing were both within specification, indicating the sensor was providing accurate glucose readings.No malfunction or product deficiency was identified.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A low readings issue was reported with the abbott diabetes care (adc) device.A customer reported receiving an unspecified low scan on the sensor when compared to an unspecified result obtained from a competitor meter.The customer experienced symptoms described as "disorientation, sticky lips, thirst, nausea, seizure, and a loss of consciousness", and was unable to self-treat.The customer had contact with a healthcare professional who administered drip administration with saline for the diagnosis of hyperglycemia.No further information was reported.There was no report of death or permanent impairment associated with this event.
|
|