• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) VICTORY XL DR; IMPLANTABLE PACEMAKER PULSE-GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) VICTORY XL DR; IMPLANTABLE PACEMAKER PULSE-GENERATOR Back to Search Results
Model Number 5816
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/13/2023
Event Type  malfunction  
Event Description
It was reported the patient presented for a routine generator exchange.During the procedure, the right ventricular lead could not be inserted into the header of the new pacemaker.The pacemaker was replaced to complete the procedure.There were no patient consequences.
 
Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.The reported event of not possible to fully insert the lead into the device v-connector was not confirmed.The pacemaker was returned for analysis.Analysis performed indicated that test leads could be fully inserted into both atrial and ventricular connector ports.All connector port dimensions were normal.Lead insertion force tests were normal.No anomalies were found.The device had normal device characteristics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VICTORY XL DR
Type of Device
IMPLANTABLE PACEMAKER PULSE-GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17222608
MDR Text Key318494369
Report Number2017865-2023-24579
Device Sequence Number1
Product Code DXY
UDI-Device Identifier05414734009553
UDI-Public05414734009553
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2024
Device Model Number5816
Device Catalogue Number5816
Device Lot NumberP000154885
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-