Catalog Number FVM12120 |
Device Problem
Misfire (2532)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/06/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during a stent graft placement procedure in the distal arteria femoralis superficialis, the stent allegedly could not be deployed more than 1.5cm.It was further reported that the delivery system blocked completely.Reportedly, the stent was removed from the patient with the delivery system.The procedure was completed using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the fluency plus endovascular stent graft products that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft products are identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.H10: d4 (expiry date: 07/2025).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the fluency plus endovascular stent graft products that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft products are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the stent delivery system was returned for evaluation and the stent graft was found partially deployed which leads to confirmed results for partial deployment.It was reported that a 10f introducer/0.035" guidewire were used for access, device was flushed, the vessel was neither calcified nor tortuous and the proximal part of the device was placed in the straight segment of the lumen before deployment attempt.Based on the provided information and the evaluation of the returned sample, the investigation is confirmed for partial deployment.The intended placement of the device to treat aneurysm represent off-label use.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.Regarding preparation of the device the instructions for use state that 'prior to loading the vascular system over a guide wire, both ports must be flushed with sterile saline.Flushing these lumens will also facilitate stent graft deployment'.Regarding the anatomy of the placement site the instructions for use states: 'prior to stent graft deployment, ensure that the proximal stent graft end is positioned in a straight section of the lumen to reduce the risk of increased deployment forces and possible failure to deploy'.Regarding accessories the instructions for use states: 'a super stiff guide wire (0.035 in.) is advanced from a femoral artery puncture site.Use an introducer sheath for the implant procedure'; the packaging pictograms indicate an introducer size of 10f and a 0.035" guidewire.The instructions for use states that "the safety and effectiveness of the device for use in the treatment of aneurysms and pseudoaneurysms have not been established".H10: d4 (expiration date: 07/2025), g3 h11: h6 (result, conclusion) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during a stent graft placement procedure in the distal arteria femoralis superficialis, the stent allegedly could not be deployed more than 1.5cm.It was further reported that the delivery system blocked completely.Reportedly, the stent was removed from the patient with the delivery system.The procedure was completed using another device.There was no reported patient injury.
|
|