Device evaluation summary: the reported instrument failing quality control was verified during service.The service technician found the drop times were out of specification.The issue was resolved by adjusting the drop times to the proper settings.The service technician noted that the error printout showed several cartridge sensor errors and both x and y axis motor stalls, the adu power cable was worn and the x axis home position was out of tolerance.The issues were resolved by replacing both drive belts, the sensors and the adu power cable, and recalibrating the switch.Post-repair testing was performed per specifications.Conclusion: complaint confirmed for the hms plus instrument failing quality control (qc).No patient/clinical safety issues reported.Trends for issues with this product are reviewed at quarterly quality meetings.The hms plus software constantly monitors for software and/or hardware faults.When one is found, an error is displayed and typically addressed by the operator.This regulatory report is being submitted as part of a retrospective review and remediation per d00953163 as part of a capa action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|