Model Number WA53005A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported to olympus that a telescope had partial eyepiece damage.It is unknown when the reported issue was found.There was no patient injury or adverse event reported to olympus.This medwatch is being submitted for the reportable malfunction found during evaluation.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and repair.During the evaluation, the reported issue was confirmed, the eyepiece was found damaged.The image failure was also traced back to cover glass adhesion.The investigation is still in progress; therefore, the root cause of the reported defect cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.In general, the customer is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.Olympus will continue to monitor the field performance of this device.Additional pma/510(k): k923982/ k950076.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to e1 and g2.Correction: corrected e1 to translate to english.Checked "other" to add the country japan.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the event occurred due to the user improperly handling and applying excessive force to the subject device.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE
Type of Device
RIGID SCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17241542
MDR Text Key318565302
Report Number9610773-2023-01777
Device Sequence Number1
Product Code EWY
UDI-Device Identifier04042761052764
UDI-Public04042761052764
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K912362
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA53005A
Device Catalogue NumberWA53005A
Device Lot Number517387
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2023
Was the Report Sent to FDA? No
Date Manufacturer Received09/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-