• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC FABRICATION S.A.S. STREAMLINE TEMPORARY PACING LEADS; ELECTRODE, PACEMAKER, TEMPORARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC FABRICATION S.A.S. STREAMLINE TEMPORARY PACING LEADS; ELECTRODE, PACEMAKER, TEMPORARY Back to Search Results
Model Number 6492F
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2023
Event Type  malfunction  
Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that on the same day post implant of this temporary pacing lead, the lead broke.The lead broke when an attempt was made to connect it to the temporary pacemaker.It was reported that the lead was not connected to the temporary pacemaker and the leads were left coiled and taped on the patient.A new lead was not inserted.It was reported that patient anatomy did not contribute to the pacing lead breakage.The broken portion that was in the patient was retrieved.It was noted that the lead was explanted 10 days post implant.It is stated that staff are experienced in the use of this pacing lead.No intervention or adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STREAMLINE TEMPORARY PACING LEADS
Type of Device
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
MEDTRONIC FABRICATION S.A.S.
103 route d anor
fourmies nord 59610
FR  59610
Manufacturer (Section G)
MEDTRONIC FABRICATION S.A.S.
103 route d anor
fourmies nord 59610
FR   59610
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17245831
MDR Text Key318559551
Report Number9611350-2023-00005
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K190716
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6492F
Device Catalogue Number6492F
Device Lot NumberBBD299206F
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/05/2023
Initial Date FDA Received07/03/2023
Date Device Manufactured11/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-