• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH ADAPTER FOR TRIAL HEAD, S/-4, TAPER 12/14; HIP INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH ADAPTER FOR TRIAL HEAD, S/-4, TAPER 12/14; HIP INSTRUMENTS Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/05/2023
Event Type  malfunction  
Event Description
It was reported that a small crack in the lower part of the item was detected.No certain cause of the fracture.No patient consequences.This event is related to a malfunction that could potentially lead to a sterility issue.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Manufacturer Narrative
(b)(4).G2-foreign- canada.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Product was returned for investigation.The instrument shows damages in the form of scratches and dents.Moreover, one side of the instrument is fractured.The new available information does not change the outcome of the investigation.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations.Device is used for treatment.Medical records were not provided.However, with the available information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADAPTER FOR TRIAL HEAD, S/-4, TAPER 12/14
Type of Device
HIP INSTRUMENTS
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key17246175
MDR Text Key318343643
Report Number0009613350-2023-00253
Device Sequence Number1
Product Code HSD
UDI-Device Identifier00889024278691
UDI-Public(01)00889024278691(11)180611(10)4502297895
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 08/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number01.00189.145
Device Lot Number4502297895
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/11/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-