Catalog Number 2C8931 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/09/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that there was no inner valve on one of the ports of a clearlink system continu-flo solution set.The manifold connections came loose event after tightening.The issue occurred after priming.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.Visual inspection was performed and the inner valve was observed missing.Under water pressure testing was performed and a leak was observed at the y-site.The reported condition was verified.The condition cause of the condition was due a supplier manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|