BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO MULTI-PORT MANIFOLD; STOPCOCK, I.V. SET Back to Search Results
Catalog Number 2C8931
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/09/2023
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was no inner valve on one of the ports of a clearlink system continu-flo solution set.The manifold connections came loose event after tightening.The issue occurred after priming.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the device was received for evaluation.Visual inspection was performed and the inner valve was observed missing.Under water pressure testing was performed and a leak was observed at the y-site.The reported condition was verified.The condition cause of the condition was due a supplier manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO MULTI-PORT MANIFOLD
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17249416
MDR Text Key318345159
Report Number1416980-2023-03307
Device Sequence Number1
Product Code FMG
UDI-Device Identifier00085412071725
UDI-Public(01)00085412071725
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K932512
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8931
Device Lot NumberDR22J11048
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/09/2023
Initial Date FDA Received07/03/2023
Supplement Dates Manufacturer Received08/28/2023
Supplement Dates FDA Received09/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/13/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NA.
-
-