Mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - investigation of the returned unit showed that the lock had both rotational and lateral movement and a residue buildup was present.Upon disassembly, it was observed that the index knob and the lock needed new components added to replace worn internal parts.To resolve all issues, the unit was machined to have heli-coils added to large starburst threads, new components were added to replace worn internal parts and general maintenance and cleaning were performed.Root cause - complaint was confirmed via inspection of the unit.There was movement in the lock and the index knob and lock required replacement of worn components.The probable root cause is routine wear or improper handling of the unit.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|