Model Number 72404257 |
Device Problems
Deflation Problem (1149); Inflation Problem (1310); Mechanical Problem (1384); Device Contamination with Chemical or Other Material (2944); Adverse Event Without Identified Device or Use Problem (2993); Air/Gas in Device (4062)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/13/2023 |
Event Type
Injury
|
Event Description
|
It was reported that this inflatable penile prosthesis (ipp) was not functioning as intended.A surgical procedure was performed during which it was noted that there was air in the pump, the device did not inflate or deflate, and there was foreign material present the reservoir.The medical staff removed the foreign material and withdrew the fluid from the reservoir; however, the pump still did not work properly.All components from the existing device were removed and replaced with a new ipp.There were no patient complications reported.
|
|
Manufacturer Narrative
|
Upon receipt at our quality assurance laboratory, this inflatable penile prosthesis (ipp) underwent a thorough analysis.Both cylinders were visually inspected, and leak tested.Visual examination revealed that both cylinders had wear at fold in the middle of their body and their kink resistant tubings (krt) were worn to the filament.No leaks were identified in the cylinders.The pump was visually inspected, leak tested, and functionally tested.During visual test, it was noted that the krt was worn to the filament, cut into two pieces and had kinked/twisted in the tubing.No leaks were identified; however, the pump did not pass the functional test.The reservoir was visually inspected, and leak tested.No damage or abnormalities were noted, no leaks were identified in the reservoir.Based on the information available and analysis results, the pump not passing the functional test could have caused or contributed to the reported clinical observations; therefore, a conclusion code of caused traced to component failure was assigned to this investigation.
|
|
Event Description
|
It was reported that the patient could not get this inflatable penile prosthesis (ipp) to work.A surgical procedure was performed during which it was noted that there was air in the pump, the device did not inflate or deflate, and there was a foreign material present the reservoir.The medical staff removed the foreign material and withdrew the fluid from the reservoir; however, the pump still did not work properly.Therefore, all components from the existing device were removed and replaced with a new ipp.There were no patient compo cations reported.
|
|
Manufacturer Narrative
|
Upon receipt at our quality assurance laboratory, this inflatable penile prosthesis (ipp) underwent a thorough analysis.Both cylinders were visually inspected, and leak tested.Visual examination revealed that both cylinders had wear at fold in the middle of their body and their kink resistant tubings (krt) were worn to the filament.No leaks were identified in the cylinders.The pump was visually inspected, leak tested, and functionally tested.During visual test, it was noted that the krt was worn to the filament, cut into two pieces and had kinked/twisted in the tubing; in addition, the deflation ball was misaligned.No leaks were identified; however, the pump did not pass the functional test.The reservoir was visually inspected, and leak tested.No damage or abnormalities were noted, no leaks were identified in the reservoir.Based on the information available and analysis results, the pump not passing the functional test could have caused or contributed to the reported clinical observations.
|
|
Event Description
|
It was reported that the patient could not get this inflatable penile prosthesis (ipp) to work.A surgical procedure was performed during which it was noted that there was air in the pump, the device did not inflate or deflate, and there was a foreign material present the reservoir.The medical staff removed the foreign material and withdrew the fluid from the reservoir; however, the pump still did not work properly.Therefore, all components from the existing device were removed and replaced with a new ipp.There were no patient complications reported.
|
|
Search Alerts/Recalls
|
|