Catalog Number 110003450 |
Device Problems
Fracture (1260); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/06/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-01481; 0001825034-2023-01482.Italy.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
|
|
Event Description
|
It was reported that during the procedure, the inserter broke upon impaction.A new instrument was opened and used to complete the procedure.Attempts have been made and no further information is available.
|
|
Event Description
|
No additional event information to report at this time.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified damage and a broken tip.The complaint is confirmed based on the provided pictures.Dhr was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified there are impact marks near the threads and to the strike plate.The threads have been damaged so no further analysis could be performed.The complaint is confirmed based on the evaluation of the returned device.Root cause remains unchanged.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|