It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited loss of capture (loc) and high capture thresholds on the right ventricular (rv) and left ventricular (lv) channels.The lv lead is a non-boston scientific product.It was noted that the capture on the right atrial (ra) channel was also in question.The ra channel also exhibited farfield oversensing of the ventricle signals.The ra channel exhibited a drop in amplitude and intermittent undersensing, as well.Additionally, the rv and ra channels' impedances have been gradually increasing but remain within range.The lv channel exhibited gradually decreasing impedance but also remain within range.It was noted that the pacing impedances on all three channels have been changing over the past 3 to 4 weeks.It was noted that the patient's respiratory rate had decreased around the same time.Additionally, this patient fell and was feeling poorly for a couple of weeks.It was noted that the patient was having a metabolic crisis around this time.The device remains in use.No adverse patient effects were reported.
|