ALCON LABORATORIES IRELAND LTD. CLAREON ASPHERIC UV ABSORBING IOL WITH AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number CCA0T0
Device Problem Particulates (1451)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A non healthcare professional reported that, the physician noticed what he calls polychromatic sheen on some of the lenses.The physician also added that he was seeing that mostly with specific model of the company lenses.Additionally the physician stated that he noticed this with lenses in the past at another location as well.Additional information was received stating that, the physician noticed it intraoperative - leaving lens implanted.The physician have seen the same thing over the last month in the company lenses as well.The physician also added that he saw this with both the pre-loaded and regular company lenses.None of the patients have any issues with visual acuity.Additionally, the physician also stated that, he did not see this before in months he was using similar company lenses.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The product was not returned for analysis.Only photo was returned.Returned photo shows, what appears to be implanted iol.There are reflections of theatre lights on iol optic.The origin of the reported complaint cannot be confirmed from the returned photo.Based on our observation of the attached photo, there are reflections of theatre lights on implanted iol optic.A final root cause cannot be determined based on available information.Based on the results from the product history record, the products met release criteria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC UV ABSORBING IOL WITH AUTONOME DELIVERY SYSTEM
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17266731
MDR Text Key318463429
Report Number9612169-2023-00514
Device Sequence Number1
Product Code HQL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P190018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCCA0T0
Device Lot Number25623878
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/08/2023
Initial Date FDA Received07/06/2023
Supplement Dates Manufacturer Received08/30/2023
Supplement Dates FDA Received09/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/05/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-